Irradiation of lung and esophagus tumors: A comparison of dose distributions calculated by anisotropic analytical algorithm and pencil beam convolution algorithm, a retrospective dosimetric study

Amankwaa-Frempong, Emmanuel
Vernimmen, Frederik
Blay, Samuel
Ezhilalan, Ramalingam
Advisors
Editors
Other Contributors
Departments
Date
2014-04-09
Date Submitted
Keywords
CANCER
RADIOTHERAPY
Other Subjects
Subject Mesh
Planned Date
Start Date
Collaborators
Principal Investigators
Alternative Titles
Abstract
Purpose: The aim of this study was to evaluate dosimetric differences between pencil beam convolution (PBC) algorithm and anisotropic analytical algorithm (AAA) calculations in patients with lung and esophageal cancers. Methods: The existing plans calculated with PBC for 60 patients treated in 2012 were recalculated with AAA maintaining the same beam geometry and dose monitor units. For these plans, dose prescription ranges were 41.4 Gy to 56.0 Gy for esophageal cancers and 50.0 Gy to 64.0 Gy for lung cancers. Dosimetric variables were the 95% PTV coverage, mean PTV dose, maximum spinal cord dose, lung V5Gy, and lung V20Gy. Results: The 95% PTV coverage’s for both lung and esophageal tumors were reduced when recalculated with AAA. Maximum spinal cord doses for lung cancer patients were reduced by 0.7 Gy and by 0.3 Gy for esophageal cancer patients on AAA. On the other hand, lung V5Gy had 3.5% increase for both lung and esophageal cancer patients on AAA, whereas lung V20Gy increased by 1.5% also on AAA for esophageal cancer patients. Conclusion: These clinical results confirm the differences between AAA and PBC algorithms as observed in phantom dosimetric studies, and give an indication of the clinical implications of changing from one calculation algorithm to another.
Language
en
ISSN
23304049
eISSN
ISBN
DOI
10.14319/ijcto.0202.10
PMID
PMCID
Sponsorships
Funding Sources
Funding Amounts
Grant Identifiers
Methodology
Duration
Ethical Approval