dc.identifier.citationThe Neopuff's PEEP valve is flow sensitive. 2011, 100 (3):360-3 Acta Paediatr.en
dc.identifier.issn1651-2227
dc.identifier.pmid21070358
dc.identifier.doi10.1111/j.1651-2227.2010.02086.x
dc.identifier.urihttp://hdl.handle.net/10147/135510
dc.description.abstractThe current recommendation in setting up the Neopuff is to use a gas flow of 5-15 L/min. We investigated if the sensitivity of the positive end expiratory pressure (PEEP) valve varies at different flow rates within this range.
dc.description.abstractFive Neopuffs were set up to provide a PEEP of 5 cm H(2) O. The number of clockwise revolutions to complete occlusion of the PEEP valve and the mean and range of pressures at each quarter clockwise revolution were recorded at gas flow rates between 5 and 15 L/min. Results:  At 5, 10 and 15 L/min, 0.5, 1.7 and 3.4 full clockwise rotations were required to completely occlude the PEEP valve, and pressures rose from 5 to 11.4, 18.4 and 21.5 cm H(2) O, respectively. At a flow rate of 5 L/min, half a rotation of the PEEP dial resulted in a rise in PEEP from 5 to 11.4cm H(2) O. At 10 L/min, half a rotation resulted in a rise from 5 to 7.7cm H(2) O, and at 15 L/min PEEP rose from 5 to 6.8cm H(2) O.
dc.description.abstractUsers of the Neopuff should be aware that the PEEP valve is more sensitive at lower flow rates and that half a rotation of the dial at 5 L/min gas flow can more than double the PEEP.
dc.language.isoenen
dc.relation.urlhttp://www.ncbi.nlm.nih.gov/pubmed/21070358en
dc.subject.meshEquipment Design
dc.subject.meshEquipment Safety
dc.subject.meshHumans
dc.subject.meshInfant, Newborn
dc.subject.meshManometry
dc.subject.meshPositive-Pressure Respiration
dc.titleThe Neopuff's PEEP valve is flow sensitive.en
dc.typeArticleen
dc.contributor.departmentDepartment of Neonatology, Cork University Maternity Hospital, Cork, Ireland.en
dc.identifier.journalActa paediatrica (Oslo, Norway : 1992)en
dc.description.provinceMunster
html.description.abstractThe current recommendation in setting up the Neopuff is to use a gas flow of 5-15 L/min. We investigated if the sensitivity of the positive end expiratory pressure (PEEP) valve varies at different flow rates within this range.
html.description.abstractFive Neopuffs were set up to provide a PEEP of 5 cm H(2) O. The number of clockwise revolutions to complete occlusion of the PEEP valve and the mean and range of pressures at each quarter clockwise revolution were recorded at gas flow rates between 5 and 15 L/min. Results:  At 5, 10 and 15 L/min, 0.5, 1.7 and 3.4 full clockwise rotations were required to completely occlude the PEEP valve, and pressures rose from 5 to 11.4, 18.4 and 21.5 cm H(2) O, respectively. At a flow rate of 5 L/min, half a rotation of the PEEP dial resulted in a rise in PEEP from 5 to 11.4cm H(2) O. At 10 L/min, half a rotation resulted in a rise from 5 to 7.7cm H(2) O, and at 15 L/min PEEP rose from 5 to 6.8cm H(2) O.
html.description.abstractUsers of the Neopuff should be aware that the PEEP valve is more sensitive at lower flow rates and that half a rotation of the dial at 5 L/min gas flow can more than double the PEEP.