dc.description.abstractSummary: A 43-year-old woman attended for stent assisted coiling. A Neuroform 30 x 4.5 mm stent had been successfully placed over the left periophthalmic aneurysm. During the coiling the first coil migrated through the crowns in the stent, lodging at the MCA bifurcation. We believe that the coil herniated through the overlying stent due to the carotid siphon curvature and the open cell design. Furthermore the distal markers of the stent impeded coil extraction with a MERCI device.
dc.language.isoenen
dc.titleCoil Migration through a Neuroform 3 Stent during Endovascular Coiling. A Case Report.en
dc.typeArticleen
dc.contributor.departmentDepartment of Neuroradiology, Beaumont Hospital; Dublin, Ireland - alanohare@yahoo.com.en
dc.identifier.journalInterventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciencesen
dc.description.provinceLeinster
html.description.abstractSummary: A 43-year-old woman attended for stent assisted coiling. A Neuroform 30 x 4.5 mm stent had been successfully placed over the left periophthalmic aneurysm. During the coiling the first coil migrated through the crowns in the stent, lodging at the MCA bifurcation. We believe that the coil herniated through the overlying stent due to the carotid siphon curvature and the open cell design. Furthermore the distal markers of the stent impeded coil extraction with a MERCI device.


This item appears in the following Collection(s)

Show simple item record