A new barbed device for repair of flexor tendons.

Hdl Handle:
http://hdl.handle.net/10147/207969
Title:
A new barbed device for repair of flexor tendons.
Authors:
Hirpara, K M; Sullivan, P J; O'Sullivan, M E
Affiliation:
Department of Trauma and Orthopaedics, Waterford Regional Hospital, Ardkeen,, Dunmore Road, Waterford, Ireland. kie_hirpara@hotmail.com
Citation:
J Bone Joint Surg Br. 2010 Aug;92(8):1165-70.
Journal:
The Journal of bone and joint surgery. British volume
Issue Date:
1-Feb-2012
URI:
http://hdl.handle.net/10147/207969
DOI:
10.1302/0301-620X.92B8.23029
PubMed ID:
20675766
Abstract:
We split 100 porcine flexor tendons into five groups of 20 tendons for repair. Three groups were repaired using the Pennington modified Kessler technique, the cruciate or the Savage technique, one using one new device per tendon and the other with two new devices per tendon. Half of the tendons received supplemental circumferential Silfverskiold type B cross-stitch. The repairs were loaded to failure and a record made of their bulk, the force required to produce a 3 mm gap, the maximum force applied before failure and the stiffness. When only one device was used repairs were equivalent to the Pennington modified Kessler for all parameters except the force to produce a 3 mm gap when supplemented with a circumferential repair, which was equivalent to the cruciate. When two devices were used the repair strength was equivalent to the cruciate repair, and when the two-device repair was supplemented with a circumferential suture the force to produce a 3 mm gap was equivalent to that of the Savage six-strand technique.
Language:
eng
MeSH:
Animals; Elasticity; Equipment Design; Stress, Mechanical; Sus scrofa; Suture Techniques/*instrumentation; Tendon Injuries/*surgery; Tendons/physiopathology
ISSN:
0301-620X (Print); 0301-620X (Linking)

Full metadata record

DC FieldValue Language
dc.contributor.authorHirpara, K Men_GB
dc.contributor.authorSullivan, P Jen_GB
dc.contributor.authorO'Sullivan, M Een_GB
dc.date.accessioned2012-02-01T10:52:52Z-
dc.date.available2012-02-01T10:52:52Z-
dc.date.issued2012-02-01T10:52:52Z-
dc.identifier.citationJ Bone Joint Surg Br. 2010 Aug;92(8):1165-70.en_GB
dc.identifier.issn0301-620X (Print)en_GB
dc.identifier.issn0301-620X (Linking)en_GB
dc.identifier.pmid20675766en_GB
dc.identifier.doi10.1302/0301-620X.92B8.23029en_GB
dc.identifier.urihttp://hdl.handle.net/10147/207969-
dc.description.abstractWe split 100 porcine flexor tendons into five groups of 20 tendons for repair. Three groups were repaired using the Pennington modified Kessler technique, the cruciate or the Savage technique, one using one new device per tendon and the other with two new devices per tendon. Half of the tendons received supplemental circumferential Silfverskiold type B cross-stitch. The repairs were loaded to failure and a record made of their bulk, the force required to produce a 3 mm gap, the maximum force applied before failure and the stiffness. When only one device was used repairs were equivalent to the Pennington modified Kessler for all parameters except the force to produce a 3 mm gap when supplemented with a circumferential repair, which was equivalent to the cruciate. When two devices were used the repair strength was equivalent to the cruciate repair, and when the two-device repair was supplemented with a circumferential suture the force to produce a 3 mm gap was equivalent to that of the Savage six-strand technique.en_GB
dc.language.isoengen_GB
dc.subject.meshAnimalsen_GB
dc.subject.meshElasticityen_GB
dc.subject.meshEquipment Designen_GB
dc.subject.meshStress, Mechanicalen_GB
dc.subject.meshSus scrofaen_GB
dc.subject.meshSuture Techniques/*instrumentationen_GB
dc.subject.meshTendon Injuries/*surgeryen_GB
dc.subject.meshTendons/physiopathologyen_GB
dc.titleA new barbed device for repair of flexor tendons.en_GB