Coil Migration through a Neuroform 3 Stent during Endovascular Coiling. A Case Report.

Hdl Handle:
http://hdl.handle.net/10147/126565
Title:
Coil Migration through a Neuroform 3 Stent during Endovascular Coiling. A Case Report.
Authors:
O'Hare, A; Thornton, J; Brennan, P
Affiliation:
Department of Neuroradiology, Beaumont Hospital; Dublin, Ireland - alanohare@yahoo.com.
Citation:
Coil Migration through a Neuroform 3 Stent during Endovascular Coiling. A Case Report. 2009, 15 (2):219-22 Interv Neuroradiol
Journal:
Interventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciences
Issue Date:
29-Jul-2009
URI:
http://hdl.handle.net/10147/126565
PubMed ID:
20465903
Abstract:
Summary: A 43-year-old woman attended for stent assisted coiling. A Neuroform 30 x 4.5 mm stent had been successfully placed over the left periophthalmic aneurysm. During the coiling the first coil migrated through the crowns in the stent, lodging at the MCA bifurcation. We believe that the coil herniated through the overlying stent due to the carotid siphon curvature and the open cell design. Furthermore the distal markers of the stent impeded coil extraction with a MERCI device.
Item Type:
Article
Language:
en
ISSN:
1591-0199

Full metadata record

DC FieldValue Language
dc.contributor.authorO'Hare, Aen
dc.contributor.authorThornton, Jen
dc.contributor.authorBrennan, Pen
dc.date.accessioned2011-03-31T12:44:54Z-
dc.date.available2011-03-31T12:44:54Z-
dc.date.issued2009-07-29-
dc.identifier.citationCoil Migration through a Neuroform 3 Stent during Endovascular Coiling. A Case Report. 2009, 15 (2):219-22 Interv Neuroradiolen
dc.identifier.issn1591-0199-
dc.identifier.pmid20465903-
dc.identifier.urihttp://hdl.handle.net/10147/126565-
dc.description.abstractSummary: A 43-year-old woman attended for stent assisted coiling. A Neuroform 30 x 4.5 mm stent had been successfully placed over the left periophthalmic aneurysm. During the coiling the first coil migrated through the crowns in the stent, lodging at the MCA bifurcation. We believe that the coil herniated through the overlying stent due to the carotid siphon curvature and the open cell design. Furthermore the distal markers of the stent impeded coil extraction with a MERCI device.-
dc.language.isoenen
dc.titleCoil Migration through a Neuroform 3 Stent during Endovascular Coiling. A Case Report.en
dc.typeArticleen
dc.contributor.departmentDepartment of Neuroradiology, Beaumont Hospital; Dublin, Ireland - alanohare@yahoo.com.en
dc.identifier.journalInterventional neuroradiology : journal of peritherapeutic neuroradiology, surgical procedures and related neurosciencesen
dc.description.provinceLeinster-

Related articles on PubMed

All Items in Lenus, The Irish Health Repository are protected by copyright, with all rights reserved, unless otherwise indicated.